SELFOC® Rod Lens

Applications

 

  Small diameter rigid endoscopes

 

The SELFOC® Rod Lens is designed as a relay lens for small diameter rigid endoscopes. It is usually used in conjunction with a SELFOC® Imaging Lens (objective lens) to form a complete system. Cover glass and stainless steel tube are available as options.

Rod Lens Ordering Info

 
Technical Charts
Optical Properties and Dimensions

Diam.(mm)Pitch Dimensions (mm)Optical Properties (@550 nm)
SRL LengthILW LengthCVG LengthTotal LengthSS Tube O.D.SRLN.A.View Angle (2θ)Chromatic Aberration (μm)
SRLILWTotal
0.35114.9
+/- 0.5
0.96
+/-0.10
0.33
+0/-0.03
16.2
+/- 0.7
0.60
+0/-0.03
0.11913.5°354984
229.8
+/- 1.0
31.1
+/- 1.2
71120
344.7
+/- 1.5
46.0
+/- 1.7
106155
459.6
+/- 2.0
60.9
+/- 2.2
142191
0.5130.2
+/- 1.0
1.39
+/-0.13
0.33
+0/-0.03
31.9
+/- 1.2
0.75
+0/-0.03
0.0849.5°7251123
260.4
+/- 2.0
62.1
+/- 2.2
144195
390.6
+/- 3.0
92.3
+/- 3.2
215266
4120.8
+/- 4.0
122.5
+/- 4.2
288339
0.7144.7 +/- 1.42.00
+/-0.20
0.35
+0/-0.03
47.1 +/- 1.70.95
+0/-0.03
0.0799.0°10682188
289.4
+/- 2.8
91.8
+/- 3.1
212294
3134.1
+/- 4.2
136.5
+/- 4.5
318400
4178.8
+/- 5.6
181.2
+/- 5.9
424506
1144.7
+/- 1.4
2.95
+/-0.30
0.50
+/-0.03
48.2
+/- 1.8
1.25
+/-0.03
0.11313.0°106102208
289.4
+/- 2.8
92.9
+/- 3.2
212314
3134.1
+/- 4.2
137.6
+/- 4.6
318420
4178.8
+/- 5.6
182.3
+/- 6.0
424526
1.3167.0
+/- 2.1
3.96
+/-0.35
0.65
+0/-0.03
71.6 +/- 2.51.55
+0/-0.03
0.09811.0°159144303
2134.0
+/- 4.2
138.6
+/- 4.6
318462
3201.0
+/- 6.3
205.6
+/- 6.7
477621
21100.5
+/- 3.2
6.54
+/-0.60
1.00
+0/-0.03
108.1
+/- 3.9
2.25
+/-0.02
0.111.5°239261500
2201.0
+/- 6.4
208.6
+/- 7.1
477738
3301.5
+/- 9.6
309.1
+/-10.3
716977
2.71140.7
+/- 4.4
9.34
+/-0.80
1.35
+0/-0.03
151.4
+/- 5.3
2.95
+/-0.02
0.10711.0°302314616
2281.4 +/- 8.8292.1 +/- 9.7604918

Specification Notes:

  • For all SRL, diameter tolerance for all SRL is +0/-0.05 mm, and refractive index on axis is 1.610 @ 550 nm.
  • Material for cover glass is F-2 with refractive index of 1.62 @ 550 nm.
  • Longitudinal chromatic aberration from C-line (656.3 nm) to F-line (486.1 mm) is approx. 0.23% per pitch for all SRL.

 

SELFOC® MicroLens – Instructions

SELFOCKey to optical parameters: (all units in millimeters unless otherwise stated)

λ Wavelength of incident light in microns (>0.55 mm)
L1 Object distance (from object point to lens’ front surface)
L2 Image distance (from lens’ back surface to image point)
N0 On-axis refractive index of SELFOC® lens
Index gradient constant (mm-1)
Z Lens length
EFL Effective focal length (from rear primary plane to rear focal plane)
BFL Back focal length (from rear lens surface to rear focal plane)
MT Transverse magnification
θ+ Maximum angle from object above axis
θ Maximum angle from object below axis
Hm Maximum object height
Ls Distance from lens surface to aperture stop

Steps for using the SELFOC® MICROLENS Tables:

  1. If the object distance for your application is known, click the sheet tab entitled “Obj. Distance”. If the desired magnification is known, click the sheet tab entitled “Magnification”.
  2. Enter the required data in the colored data cells. As you enter numeric values, the SELFOC® lens parameters such as N0, √A, and EFL will be recalculated in the lens table.
  3. Adjust the Pitch in small increments and observe how the optical parameters are altered. Recall that 2πP=Z√A.

 
Physics of SELFOC

The Gradient Constant

The SELFOC lens utilizes a radial index gradient. The index of refraction is highest in the center of the lens and decreases with radial distance from the axis. The following equation describes the refractive index distribution of a SELFOC lens:

Equation 1: 
N(r) = N0(1 – ((√A2)/ 2 * r2)

This equation shows that the index falls quadratically as a function of radial distance. The resulting parabolic index distribution has a steepness that is determined by the value of the gradient constant, √A. Although the value of this parameter must be determined through indirect measurement techniques, it is a characterization of the lens’ optical performance. How rapidly rays will converge to a point for any particular wavelength depends on the gradient constant. The dependence of √A and N0, on wavelength is described by the dispersion equations listed at the end of this product guide. Note that different dispersion equations apply to different lens diameters and numerical apertures.

Lens Length & Pitch

In a SELFOC lens, rays follow sinusoidal paths until reaching the back surface of the lens. A light ray that has traversed one pitch has traversed one cycle of the sinusoidal wave that characterizes that lens. Viewed in this way, the pitch is the spatial frequency of the ray trajectory.

Equation 2:
2πP=(√A)*Z

The above equation relates the pitch (P) to the mechanical length of the lens (Z) and the gradient constant. The figure below illustrates different ray trajectories for lenses of various pitch. Notice how an image may be formed on the back surface of the lens if the pitch is chosen appropriately.

Paraxial Optics

In contrast to the optics of homogeneous materials, gradient-index optics involve smoothly-varying ray trajectories within the GRIN media. The paraxial (first-order) behavior of these materials is modeled by assuming sinusoidal ray paths within the lens and by allowing the quadratic term in Equation 1 to vanish in the ray-tracing calculations. All of the usual paraxial quantities may be calculated with the help of the ray-trace matrices given at the end of this product guide. The formulae for common paraxial distances have also been tabulated for quick reference.

 

Recommended Storage and Handling of Lenses

Storage
For extended periods of time, the lenses should be stored in a “dry box” environment (40%RH or less). This entails the use of a desiccant (e.g., silica gel) or a heat source to prevent humidity from leaching the lens material. This is much more critical for non-coated lenses, since AR coatings help to protect the lens surfaces from humidity. For short term storage (less than a month), the plastic box and foam packing in which the lenses are shipped will provide adequate storage.

In addition to humidity requirement, the lenses need to have sufficient spacing to avoid potential damage such as chipping and scratching from other lenses. For this reason, Go!Foton storage boxes have built-in slots in which the lenses are placed, with surrounding packaging to hold them securely in place.

Handling
After opening the lens boxes, it is important to exercise extra care in lifting the plastic shield. Particularly with smaller lenses, it is possible that they may cling to the shield and be lost during removal. Lenses should be handled with plastic tweezers, preferably those with a tapered end. Lenses should be picked up out of their individual compartments by firmly holding each by its side surface (not the ends).

Cleaning
At times it is necessary to clean the lens surfaces due to the presence of some dust or film which may impair the image. Go!Foton generally recommends the use of ethyl alcohol as a cleaning solvent. Acetone may also be used, without harm to the lens, but it should be pure enough to no leave a residue on the lens’ surface.